Study on Thermoelectrically Cooled CCD in Raman Spectrum Detection System for Gas Analysis
Release time:2018-09-26  Hits:

Indexed by: Journal paper

Document Code: 070401

First Author: 赵俊

Correspondence Author: 左都罗

Co-author: 赵俊; 左都罗; 王新兵

Journal: Laser & Optoelectronics Progress

Affiliation of Author(s): 武汉光电国家实验室

Place of Publication: Shanghai

Document Type: C

Volume: 53

Issue: 7

Page Number: 070401-1-070401-9

ISSN No.: 1006-4125

Date of Publication: 2016-07-15

Abstract: The commonly used CCD in Raman spectroscopy is sensitive to temperature,and cooling can effectively reduce its dark current noise. A cooling system is designed,and the finite element analysis software ANSYS is used to simulate the heat conduction of the heat sink and the heat exchange of gas flow in the vacuum chamber. The temperature change with the cooling time is measured at different vacuum degrees and with different buffer gas including Ar,Ne,N_2 and air. The experimental results show that better cooling effect can be realized when N_2 is used as buffer gas,and higher vacuum degree ...